View : 601 Download: 0

Fully reflective deep ultraviolet to near infrared spectrometer and entrance optics for resonance Raman spectroscopy

Title
Fully reflective deep ultraviolet to near infrared spectrometer and entrance optics for resonance Raman spectroscopy
Authors
Schulz B.Backstrom J.Budelmann D.Maeser R.Rubhausen M.Klein M.V.Schoeffel E.Mihill A.Yoon S.
Ewha Authors
윤석현
SCOPUS Author ID
윤석현scopus
Issue Date
2005
Journal Title
Review of Scientific Instruments
ISSN
0034-6748JCR Link
Citation
Review of Scientific Instruments vol. 76, no. 7
Indexed
SCI; SCIE; SCOPUS scopus
Document Type
Article
Abstract
We present the design and performance of a new triple-grating deep ultraviolet to near-infrared spectrometer. The system is fully achromatic due to the use of reflective optics. The minimization of image aberrations by using on- and off- axis parabolic mirrors as well as elliptical mirrors yields a strong stray light rejection with high resolution over a wavelength range between 165 and 1000 nm. The Raman signal is collected with a reflective entrance objective with a numerical aperture of 0.5, featuring a Cassegrain-type design. Resonance Raman studies on semiconductors and on correlated compounds, such as LaMn O3, highlight the performance of this instrument, and show diverse resonance effects between 1.96 and 5.4 eV. © 2005 American Institute of Physics.
DOI
10.1063/1.1946985
Appears in Collections:
자연과학대학 > 물리학전공 > Journal papers
Files in This Item:
There are no files associated with this item.