dc.description.abstractA column-switching high-performance liquid chromatographic (HPLC) method has been developed and validated for quantification of fluvastatin in rat plasma. Plasma samples were diluted with an equal volume of mobile phase, i.e. acetonitrile-5 mM potassium phosphate buffer (pH 6.8) (15:85, v/v), and the mixture was directly injected onto the HPLC system. The analyte was enriched in a pre-treatment column, while endogenous components were eluted to waste. The analyte was then back-flushed onto an analytical column and quantified with fluorescence detection (λ ex = 305 nm; λ em = 390 nm). The standard curve for the drug was linear in the range 0.5-100 ng mL -1 in rat plasma. The limit of quantitation for plasma was found to be 0.5 ng mL -1. This method has been fully validated and shown to be specific, accurate and precise. The method is simple and rapid because of a minimized sample preparation and appears to be useful for the pharmacokinetic study of fluvastatin. © 2006.*
dc.languageEnglish*
dc.titleColumn-switching high-performance liquid chromatographic analysis of fluvastatin in rat plasma by direct injection*
dc.typeArticle*
dc.relation.issue4*
dc.relation.volume41*
dc.relation.indexSCI*
dc.relation.indexSCIE*
dc.relation.indexSCOPUS*
dc.relation.startpage1458*
dc.relation.lastpage1462*
dc.relation.journaltitleJournal of Pharmaceutical and Biomedical Analysis*
dc.identifier.doi10.1016/j.jpba.2006.03.021*
dc.identifier.wosidWOS:000238491100059*
dc.identifier.scopusid2-s2.0-33646915183*
dc.author.googleUm S.Y.*
dc.author.googleJung S.H.*
dc.author.googleJung S.J.*
dc.author.googleKim J.I.*
dc.author.googleChung S.Y.*
dc.author.googleLee H.J.*
dc.author.googleHan S.B.*
dc.author.googleChoi S.O.*
dc.contributor.scopusid이화정(57102029300)*
dc.date.modifydate20240118154655*
Appears in Collections:
약학대학 > 약학과 > Journal papers
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

BROWSE