Large-scale performance evaluation of Accu-Chek inform II point-of-care glucose meters
Title
Large-scale performance evaluation of Accu-Chek inform II point-of-care glucose meters
Authors
Jeong T.-D.
;
Cho E.-J.
;
Ko D.-H.
;
Lee W.
;
Chun S.
;
Hong K.-S.
;
Min W.-K.
Ewha Authors
홍기숙
;
정태동
SCOPUS Author ID
홍기숙
; 정태동
Issue Date
2016
Journal Title
Scandinavian Journal of Clinical and Laboratory Investigation
ISSN
0036-5513
Citation
Scandinavian Journal of Clinical and Laboratory Investigation vol. 76, no. 8, pp. 657 - 663
Keywords
Blood glucose
;
diabetes
;
performance evaluation
;
point-of-care testing
Publisher
Taylor and Francis Ltd
Indexed
SCI; SCIE; SCOPUS
Document Type
Article
Abstract
Background: The aim of this study was to report the experience of large-scale performance evaluation of 238 Accu-Chek Inform II point-of-care (POC) glucose meters in a single medical setting. Methods: The repeatability of 238 POC devices, the within-site imprecision of 12 devices, and the linearity of 49 devices were evaluated using glucose control solutions. The glucose results of 24 POC devices and central laboratory were compared using patient samples. Results: Mean concentration of control solutions was 2.39 mmol/L for Level 1 and 16.52 mmol/L for Level 2. The pooled repeatability coefficient of variation (CV) of the 238 devices was 2.0% for Level 1 and 1.6% for Level 2. The pooled within-site imprecision CV and reproducibility CV of the 12 devices were 2.7% and 2.7% for Level 1, and 1.9%, and 1.9% for Level 2, respectively. The test results of all 49 devices were linear within analytical measurement range from 1.55–31.02 mmol/L. The correlation coefficient for individual POC devices ranged from 0.9967–0.9985. The total correlation coefficient for the 24 devices was 0.998. Conclusions: The Accu-Chek Inform II POC blood glucose meters performed well in terms of precision, linearity, and correlation evaluations. Consensus guidelines for the large-scale performance evaluations of POC devices are required. © 2016 Medisinsk Fysiologisk Forenings Forlag (MFFF).
DOI
10.1080/00365513.2016.1233575
Appears in Collections:
의과대학
>
의학과
>
Journal papers
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML
Show full item record
Find@EWHA
트윗하기
BROWSE
Communities & Collections
By Date
Authors
Titles
Subject